13. Information Provided by the Manufacturer
13.1. Each device must be accompanied by the information necessary to ensure its safe and proper use, taking into account the training and knowledge of potential users, and to identify the manufacturer.
This information consists of the instructions contained in the user manual.
To the extent possible and appropriate, the information necessary to use the device safely must appear on the device itself and/or on the packaging of each unit or, where applicable, on the commercial packaging. If it is not possible to package each unit separately, the information must appear in a leaflet accompanying one or more devices.
The packaging of each device must contain an instruction manual. An exception is made for devices in Classes I and IIa, if they can be used safely without the aid of such instructions.
a) the name or business name and address of the manufacturer. For devices imported into the Community for distribution there, the labeling, outer packaging, or instructions for use shall also include the name and address of the authorized representative if the manufacturer is not established in the Community;
b) the information strictly necessary to identify the device and the contents of the package, particularly for users;
c) where applicable, the word “STERILE”;
d) where applicable, the lot code, preceded by the word “LOT,” or the serial number;
e) where applicable, the date until which the device should be used safely, expressed as the year and month;
f) where applicable, a statement specifying that the device is for single use. The manufacturer’s indication of single use must be uniform throughout the Community;
g) in the case of a custom-made device, the words “custom-made device”;
h) if the device is intended for clinical investigations, the statement “exclusively for clinical investigations”;
i) specific storage and/or maintenance conditions;
j) specific instructions for use;
k) warnings and/or precautions to be taken.
13.2. Where applicable, this information should be presented in the form of symbols. Any symbol or identification color must comply with harmonized standards. In areas where no standards exist, the symbols and colors must be described in the documentation provided with the device.
13.3. The labeling must include the following information:
l) the year of manufacture for active devices, other than those covered by point e). This information may be included in the lot or serial number;
m) where applicable, the sterilization method;
n) in the case of a device as defined in Article 1, paragraph 4a, a statement indicating that the device incorporates, as an integral part, a substance derived from human blood.
13.4. If the intended use of the device is not obvious to the user, the manufacturer must clearly state it on the labeling and in the instructions for use.
13.5. To the extent reasonably possible, devices and detachable components must be identified, where applicable by batch, so as to enable any appropriate action to be taken to detect a potential risk associated with the devices and detachable components.
13.6. The instructions for use must include, where applicable, the following information:
a) the information referred to in section 13.3, with the exception of that listed in sections d) and e);
b) the performance characteristics referred to in item 3, as well as any adverse side effects;
c) if the device must be installed with or connected to other medical devices or equipment to function as intended, sufficient information regarding its characteristics to identify the correct devices or equipment that must be used to ensure a safe combination;
d) all information necessary to verify that the device is properly installed and can function correctly and safely, as well as instructions regarding the nature and frequency of maintenance and calibration procedures required to ensure the continued proper functioning and safety of the devices;
e) where applicable, information to help avoid certain risks associated with the implantation of the device;
f) information regarding the risks of mutual interference associated with the presence of the device during specific investigations or treatments;
g) the necessary instructions in the event of damage to the packaging ensuring sterility and, where applicable, an indication of the appropriate methods for re-sterilization;
h) if the device is intended for reuse, information regarding the appropriate procedures for reuse, including cleaning, disinfection, packaging, and, where applicable, the sterilization method if the device must be resterilized, as well as any restrictions on the number of times it may be reused. When the devices supplied must be sterilized before use, the cleaning and sterilization instructions must be such that, if properly followed, the device continues to meet the requirements of Section I. If the device bears a label indicating that it is for single use only, information on the known characteristics and technical factors known to the manufacturer that could pose a risk if the device were to be reused. If, in accordance with section 13.1, no instructions for use are required, the information must be made available to the user upon request;
i) information regarding any additional processing or handling required before the device can be used (e.g., sterilization, final assembly, etc.);
j) in the case of devices that emit radiation for medical purposes, information on the nature, type, intensity, and distribution of such radiation.
The instructions for use must also include information enabling medical personnel to inform the patient of contraindications and precautions to be taken. This information includes, in particular:
k) precautions to be taken in the event of a change in the device’s performance;
l) precautions to be taken regarding exposure, under reasonably foreseeable environmental conditions, to magnetic fields, external electrical influences, electrostatic discharges, pressure or pressure variations, acceleration, thermal ignition sources, etc.;
m) sufficient information on the medicinal product(s) that the device in question is intended to administer, including any restrictions on the choice of substances to be administered;
n) precautions to be taken against any special or unusual risks associated with the disposal of the device;
o) a medicinal product or a substance derived from human blood incorporated into the device as an integral part in accordance with Section 7.4;
p) the degree of accuracy specified for measuring devices;
q) date of publication or last revision of the instructions for use.